Gating:
Gating will allow you to view cells of interest by any combination of criteria that you choose. Gating does not change the intensity value assigned to an event as is the case for changes in voltage or compensation. It simply lets you decide which data to view and which data to ignore or discard. It is important to check that small changes in your gates don't have significant effects on your results or else your data will be prone to artifact. When you create or format a data plot (i.e. on the Data menu select "Format Histogram" or "Format Dot Plot") you can select any of the gates that you have created. This will filter the data and plot will only display those events which meet the gate criteria. Gating will not discard data unless you have requested this under "Acquisiton and Storage". Gating can subsequently be changed when you analyze your data without any loss of information.
To set up a gate you first draw a "Region" using one of the tools on the tool palette (there are four geometric shapes to choose from outlined with dotted lines). Note that the "Marker bar" (designated with an M) and the quadrant maker tool next do not define regions. They are used for statistical analysis only and can't be used to filter data like the regions/gates.
Regions which are commonly employed include:
PI for DNA content: FL2 area vs FL2 width. This window is useful for gaiting out apoptotic cells (lower left quadrant) and doublets (a separate cloud with increased FL2 width).
FSC vs SSC: This is useful for gating out RBC, myeloid cells etc, from blood or marrow.
Each region that is setup automatically defines a gate (e.g. G1 = R1). To delete regions select "Region List" under the Gates menu. To combine regions into more complex gating criteria use the "Gate List" under the Gates menu. (e.g. G5 = (R1 or R2) and R3).
Acquiring Data:
In the "Acquisition" window deselect the "Setup" button. Now when you select "Acquire" the cytometer will collect 10,000 events and automatically save a data file with the location and naming convention that you previously specified. Vortex each sample before running. Select a flow rate button on the machine (Low, Med, or High) to keep the event rate between 200 - 1000 events per second. The maximum event rates are around 4000/sec, attempts to increase the flow rates in this case will actually reduce the event rate by increasing the percentage of events that the cytometer rejects. Enter the Sample ID values if you wish these to be stored in each data file. (These can later be viewed under "Histogram Stats" for each data file).
When you finish running your samples consider saving your template for future analysis and consider saving your instrument settings. Also consider moving your data files and template to the lab computer for analysis. Print out your histograms for your notebook.
Troubleshooting:
If you get a "cytometer not ready" you probably left the cytometer on "standby". Otherwise you might be running out of sheath fluid. Check the reservoir directly or select "Status" under the Cytometer menu. Don't let the sheath fluid run dry or you'll have trouble from air bubbles clogging the cytometer.
Shut Down:
The cytometer should be left on in "Standby" mode with a tube of water on the intake nozzle when not in use. If you are the last person scheduled to use the cytometer you should follow the shutdown procedures listed in the flow lab. Basically this consists of running dilute bleach, detergent and then water through the machine each for 5 minutes and then power down both the cytometer and the computer.
Sign Out:
Record the file numbers that you used in the log book by the cytometer so the next user doesn't overwrite your data files by accident. Also sign out at the computer indicating the budget number and the time that you used the machine. It's important to do this correctly for your data files to be transferred to your Fred account.
Analyzing Data:
Stored data files may be analyzed after acquisition is complete. You may use your previous template to make matters easier but this is not necessary. From an open template select an individual histogram or data plot and select "Format Histogram (Command+F)". From the dialogue box you must change the selection from Acquisition to Analysis. Select the data file you wish to open. Repeat this process for each histogram or dataplot in the template. To increment the histograms to the next data file select them and then press (Command + ] ) or "Next data file". Statistics may also be viewed on selected histograms, and histograms may be overlayed on on another in the analysis mode.
日前,國產期刊TheInnovation獲得首個影響因子(IF=32.1),成為科睿唯安JCR綜合性期刊分類下排名僅次于《自然》(IF=64.8)和《科學》(IF=56.9)的期刊,并且這本期刊在目前......
近日,服務科學領域的全球領導者賽默飛世爾科技(以下簡稱賽默飛)宣布,在達成收購意向兩個月之后,賽默飛以28億美元、折合人民幣約190億元的價格,完成了對TheBindingSiteGroup的全現金收......
11月15日,施普林格·自然和TheLens平臺宣布結成重要的合作伙伴關系,以更深入地揭示學術研究和數據如何能通過經濟和社會成效,加速推動創新的問題解決方式。通過將科學、投資和企業領域的開放數據更好地......
一次性反應技術近年來逐漸成為生物制藥企業、CDMO公司的選擇,由于其污染風險少、清潔、驗證成本低的特點,備受業內好評。一次性技術選擇不僅僅為生物制藥企業和CDMO公司帶來一種創新方向的選擇,更重要的是......
蛋白質作為構成人體組織器官的支架和主要物質,在人體生命活動中起著重要作用。蛋白質的相互作用能產生許多效應,如形成特異底物作用通道、生成新的結合位點、失活、作用底物專一性和動力學變化等,細胞的代謝、信號......
2021年9月9日,無錫臻和生物科技有限公司(以下簡稱“臻和科技”)與美國VyantBio公司簽署TissueofOrigin?(以下簡稱“TOO?”)全球權益和ZL轉讓協議,全資收購這款唯一獲FDA......
2021年7月20日,JournalofCellularPhysiology及JournalofCellularBiochemistry同時撤回了中國學者49篇文章。從2019年開始,Journalo......
6月10日,QS教育集團正式發布了2021年世界大學排名,中國共有83所高校上榜,包括內地高校51所,港澳臺地區高校32所。中國大學的總體排名情況已經連續數年呈上升趨勢,今年再度刷新了榜單。大學排名,......
磷酸甘油酸突變酶1(PGAM1)通過其代謝活性以及與其他蛋白質(例如α平滑肌肌動蛋白(ACTA2))的相互作用,在癌癥代謝和腫瘤進展中起關鍵作用。變構調節被認為是發現針對PGAM1的高選擇性和有效抑制......
作為一種重要的植物激素,茉莉酸不僅調控植物對于機械損傷、昆蟲取食和腐生型病原菌侵害的防御反應,還參與調控諸多生長發育過程。basicHelix-Loop-Helix(bHLH)類型轉錄因子MYC2是茉......