    Lot-No.
    Ref. FR340 Expiry time: 1 year

    100 Tests (Ready to use PCR kit)  

    Zika Virus    (Real time) DOUBLE CHECK                                                    

    Only for in vitro use -

    only for research use

    -To be used by a technical person-

    Principle and use

    This amplification kit has been manufactured by Genekam Biotechnology AG, Germany to detect Zika-Virus (Double check. This is an absolute quantification assay. It tests twice the samples.

    Real time PCR is based on fluorogenic dyes. In our kit we use 2 dyes (2 probes), first Probe they are 6-Carboxy tetramethyl rhodamine (quencher) and Carboxy-fluorescein (reporter; called also FAM-Channel). Second Probe is labelled with HEX reporter (Filter for is VIC or in some machines, it is HEX)on one side and on the other side 6-Carboxy tetramethyl rhodamine (quencher). Up to 40 Ct for each dye should be taken positive. Value between 40-45 Ct for each dye should be taken as marginal positive (doubtful).
    This kit needs RNA which can be isolated from blood, serum, oral samples, liver samples, urine samples, spleen, brain tissue, vaccine, seminal samples, lymph nodes, cell cultures, tissue and any body fluid. Kindly use good methods to isolate the RNA. Kindly take common safety laboratory precautions during working  as Zika virus is highly infectious. Please use gloves during work and masks. Proceed clean and carefully otherwise you may cause contamination problems. Do not touch other objects like pens, chairs etc. during Part 1.

    IMPORTANT: we added cotton or sponge in the lid of container of the kit, to avoid damage during transportation. Please remove this cotton or sponge from the lid of each container before storage.

    Composition:

    It contains the following: (Warning: THAW THE TUBES SLOWLY: NEVER THAW IN HEATING BLOCK OR WITH HEAT FROM HAND):

    ? Tube A (2 tubes)

    ? Tube B (2 tubes)
    ? Tube Y (1 tube)

    ? Positive (+Ve) control (D1): to be stored at -20°C (1 tube): It must be used with pipette tips with filter or it should be used with separate pipettor.

    ? Negative (-Ve) Control (tube D2) (1 tube)

    Please check them before you start. Please keep all tube away from light.

    Equipments needed:

    ? Realtime PCR thermocycler

    ? Laboratory centrifuge

    ? microtubes (0.2ml)

    ? sterile Pipette-tips with and without filter (20μl, 5μl & 1μl)

    ? Pipettes (quality pipettes)

    ? Paper

    ? Pen

    ? Microtube

    ? Ice

    ? Vortexer
    ? 96 well PCR plate

    Procedure:

    ONCE AGAIN:

    VERY IMPORTANT !  PLEASE USE GLOVES ! DON’T TOUCH ANY OTHER OBJECTS, OTHERWISE THERE MAY BE RNASE CONTERMINATION DURING THIS PART.

    STEP A

    1. Kindly thaw one tube of: A, B, Y, D1 and D2. After thawing, kindly put the tubes at 4°C (as it is better). If the kit is not in use, store them at -20°C.

    2. Mark your microtubes with a sample number,  +ve Control and –ve Control. You can use 96 well microplate instead of tubes.

    3. Add 7μl of tube A to each tube.

    4. Add 10μl of B to each micro tube. Avoid to touch the wall of the microtubes.


    5. Add 1μl of Y to each tube (avoid to touch the wall of the microtubes).

    TIP: Add 7μl A + 10μl B + 1μl Y = 18μl per reaction. In case you want to run 10 reactions i.e. you need total 180μl, therefore you should mix 70μl of A + 100μl of B + 10μl of Y = 180μl from which you can take 18μl and add to each tube. This way you save time and hardware.

    6. Add 2μl of your RNA with sterile pipette-tip with filter to each micro tube according to your label except +Ve and -Ve (Avoid touching the wall).  Use every time a new pipette tip (for each sample) !

    7. Use new pipette tip with filter. Add 2μl of tube D1. This is the positive control supplied with our kit. Mix it.

    8. Use a new pipette tip. Add 2μl of –Ve (tube D2) to –Ve Control (don’t touch the wall).Mix it.

    9. Centrifuge all tubes for 20 sec. for 8000 rpm (this is not necessary but it is better).
    10. Run the program of your thermocycler as followings: Kindly check whether you have added everything correctly as the level of the volume of each micro tube must be almost the same.

    You must use quencher and reporter dye to setup your software (see FAQ) and run the following program:

    1. 60 minutes at 42°C
    10 minutes at 70°C

    2. 15 seconds at 95°C   x 45 cycles
    60 seconds at 60°C

    Before you start the PCR program, kindly check whether tubes are closed properly. Microtubes must be in contact with metal block (important!). There should be no air or lose contact with metal block of thermocycler.

    11. After step 10 is finished take out the microtubes.

    STEP B

    Once the program will be finished one can see the graphics. The negative control should run along with the bottom and positive control must give a curve in the software graphics. Use your software to analyse the results.

    Results: Usually there will be samples with two curves i.e. they are positive. There will be samples with one curve i.e they are positive. Samples without any curve are negative. The samples with two curves should have two Ct values i.e. they are double positive. The samples with one curve has only one Ct value i.e. positive samples. The samples without any curve will have no Ct values i.e. they are negative.

    If you should find any mistakes, please let us know. Thank you.

    Suggestion:
    This manual has been written specifically for beginners, hence persons with experience in PCR must use their experience to keep each step as small as possible e.g. you should calculate the amount of the needed chemicals, before starting with testing.

     

    Last update: 15--12-2016

    v1.0

    Genekam Biotechnology AG
    Duissernstr. 65a
    47058 Duisburg

    Germany

    Tel. (+49) 203 / 555858-31,-32,-33

    Fax (+49) 203 / 35 82 99

    anfrage@genekam.de

    http://www.genekam.de


    FAQ:

    1) Q: I cannot find quencher and reporter dye in my software:

    A: Many software has got the words:  FAM (as reporter) and TAM (as quencher). Therefore select both in your software.

    If your machines has only one word (for some machines only use the word FAM) you should select this one.  For HEX, you can use HEX or VIC or JOE in your Software.

    2) Precautions during the pipetting: Positive control should be pipetted with a separate pipettor and it should not be used for pipetting unknown samples or one should use pipette tips with filte


    相關文章

    細思極恐!AI能制造合成病毒了

    AI的觸角,已然從文字創作和代碼編寫,延伸到了生命底層的基因語言。斯坦福大學與Arc研究所的研究團隊,用Evo1、Evo2兩款AI基因組模型,從零“寫”出了16種自然界從未存在過的病毒,而且它們活了過......

    病毒與細菌在太空進化模式與地球不同

    美國威斯康星大學麥迪遜分校科學家開展的一項最新研究發現:在太空微重力環境中,病毒與細菌之間“生死較量”依然存在,但卻呈現出與地球不同的模式。這些發現有助于科學家設計出更高效的噬菌體療法,對抗日益嚴峻的......

    老蛋白新功能:病毒逃逸的“秘密武器”

    近日,中國農業科學院蘭州獸醫研究所科研團隊研究發現,小RNA病毒里的“老熟人”3D聚合酶,不只是病毒復制的核心,還是逃逸免疫系統的“幫兇”。相關研究成果發表在《細胞分子生物學》上。長期以來,科學家僅知......

    深海沉積物病毒多樣性和輔助代謝功能研究獲進展

    在海洋環境中,病毒是僅次于原核生物的第二大生物量成分。它們的生態意義不僅限于宿主裂解,還包括水平基因轉移和通過輔助代謝基因調節宿主代謝。這些輔助代謝基因可以編碼關鍵的代謝酶,在感染期間有效地重塑細胞過......

    抗基孔肯雅病毒中和抗體及作用機制研究獲進展

    基孔肯雅熱是由基孔肯雅病毒(CHIKV)引起、經蚊媒傳播的人畜共患急性病毒性疾病,該病毒歸類于披膜病毒科甲病毒屬。目前,已有兩款CHIKV疫苗在海外獲批,但全球范圍內仍缺乏獲批的特異性抗病毒藥物。因此......

    流行期提前、傳染性強孩子有這些癥狀速就醫

    10月9日,國家衛生健康委舉行新聞發布會。浙江省疾控中心研究員蔣健敏在會上介紹,國慶中秋假期結束,旅游返程后需重點關注三類傳染病:呼吸道傳染病、消化道傳染病和蟲媒傳染病。當前氣溫變化大,不少人因免疫系......

    研究闡明蚜蟲高效傳播作物病毒新機制

    動物的循環系統和植物的維管系統,是保障它們物質養分高效運輸和交換的重要通道,同時也是病毒系統性侵染的最為有效的路徑。為應對這種威脅,動植物均演化出了功能上高度相似的阻塞性防御機制。當病毒侵染時,動物的......

    新療法實現細菌與病毒協同抗癌

    美國哥倫比亞大學和洛克菲勒大學科學家利用細菌作為“特洛伊木馬”,繞過人體免疫系統的監控,將病毒直接運送至腫瘤內部。隨后,細菌與病毒協同作戰,對癌細胞發起強力攻擊。相關研究成果發表于最新一期《自然·生物......

    研究發現病毒蛋白抑制水稻免疫途徑新機制

    水稻作為最重要的糧食作物,為超過半數的世界人口提供主食。然而,水稻黑條矮縮病毒(SRBSDV)等病毒嚴重危害水稻生長,威脅糧食生產安全,解析病毒—水稻互作的分子機制對水稻病毒病的防控具有重要意義。近日......

    研究發現“細菌+病毒”可增強抗癌療效

    病毒在治療各類癌癥方面展現出巨大潛力,但免疫反應限制了其僅適用于體表附近腫瘤。如今,科學家證實,通過基因工程細菌包裹病毒可突破這一限制,顯著延緩小鼠體內惡性腫瘤的生長速度,這意味著將細菌與病毒結合可進......

  • <table id="4yyaw"><kbd id="4yyaw"></kbd></table>
  • <td id="4yyaw"></td>
  • 调性视频