    A. Gilson-style Pipetmen Micropipets (and similar models)


    We primarily use the Gilson micropipets in the core course labs. We have four sizes identified by the number on the round button on the plunger. The value is the maximum volume in microliters that can be transferred with that size pipet.


     

    Size indicator button on a Gilson P1000.

     Other autopipet sizes we have on hand. P1000 P200 P100 P20


    What size of micropipet is right for the job?

    RULE OF THUMB: Always select the SMALLEST size pipet that will handle the volume you wish to move to achieve the greatest accuracy. Accuracy decreases as you use unnecessarily large pipets for small volumes.

    The following table shows the useful volume ranges for each pipet type. Make special note of the min and max values.


     


     SIZE

     


     Useful Range

     


     Max Volume

     


     Min Volume


    Gilson-style P1000;
    Finnpipette 100-1000 ul


     ~200-1000 ul


     1000 ul


     200 ul


     Gilson-style P200;
    Finnpipette II 20 -200 ul


     20-200 ul


     200 ul


     20 ul


     Gilson-style P100


    10-100 ul


     100 ul


     10 ul


     Gilson-style P20;
    Finnpipette II 2 -20 ul


     ~0.5-20 ul;
    2 - 20 ul


     20 ul


     0.5 ul; 2 ul


     

    B. How to Read the Volume on the Micropipet

    Look at the front face of the pipet and you will see a window with 3 (three) digits inside. The diagram below shows the MAXIMUM value that can or should be dialed in on each size pipet. To exceed these values will put the pipet out of calibration. Beside each "window" below is the numbers place it represents. Please take the time to learn how to read them so as to avoid damaging them by dialing values out of their range.

    Note that some of the smaller volume models have one or two decimal places.



     


    Gilson-style P1000 scale window.
       P1000   P200,P100   P20   11000's  100's 10's 2,1100's 10's 1's  2 10's 1's 1 decimal  0  0  0  0  0  0


     

     

    C. How to change the load volume:

    Hold the pipetor horizontally with the plunger to your left. Change or set the load volume by turning the knurled plastic knob at the top of the hand grip. Roll it towards you to decrease the values and away from you to increase. WATCH CAREFULLY as you change the values to make sure you dont take the numbers out of range. If in doubt, review the allowable volumes for each size pipetor, and then ask for help.

     

    D. Technique

    • GRIP: There are two typical ways that people hold micropipets: the index finger grip, and the thumb grip (see figure below). Hold the pipetor like a knife in a horror flick with your index finger OR thumb on the plunger. While it's a matter of personal preference, some argue that your index finger gives you much finer control over the plunger action. As seen in the photos below, however, some micropipets are clearly designed for the thumb grip.



     


     


     Index finger grip


     Thumb grip


    • LOADING: Load a sterile tip (blue for 1000 ul pipettes; yellow for P200 and all smaller sizes including) and then reclose the tip box to maintain sterilitiy.

      • Push the plunger down slowly to the point of first resistance: this is the load volume.

      • While holding the plunger at the load volume set point, put the tip into the solution so that it is immersed just enough to cover the end (3-4 mm).

      • Slowly release the plunger to draw up the liquid making sure to keep the tip immersed. Visually inspect the load to make sure it is correct - there should be no air space in the distal end tip.

      • Performance may be improved by prewetting the tip once or twice (load and discharge a small amount) before actually loading for delivery.

    • DELIVERY: To deliver the volume, place the tip into the receiving vessel and press the plunger all the way to the bottom - this expells all the liquid and gives a little extra volume to get the last droplet out - AND THEN, WITHOUT RELEASING THE PLUNGER, withdraw the tip.

    • CHANGE TIP? Repeat as necessary using the same tip if NOT changing solutions/cultures. Use a fresh tip for every change of solution, or whenever it is prudent to maintain sterile conditions and avoid cross contamination of solutions or cultures. 

    • TIP DISCHARGE: While holding the tip over an appropriate waste receptacle, press the white tip discharge slider on the back of the grip.


    相關文章

    回顧和預測現代生物制藥為何青睞一次性生物反應器?

    一次性反應技術近年來逐漸成為生物制藥企業、CDMO公司的選擇,由于其污染風險少、清潔、驗證成本低的特點,備受業內好評。一次性技術選擇不僅僅為生物制藥企業和CDMO公司帶來一種創新方向的選擇,更重要的是......

    科學家將人工智能技術成功用于蛋白質復合物結構預測

    蛋白質作為構成人體組織器官的支架和主要物質,在人體生命活動中起著重要作用。蛋白質的相互作用能產生許多效應,如形成特異底物作用通道、生成新的結合位點、失活、作用底物專一性和動力學變化等,細胞的代謝、信號......

    發力癌癥分子病理診斷,無錫臻和全資收購TissueofOrigin?

    2021年9月9日,無錫臻和生物科技有限公司(以下簡稱“臻和科技”)與美國VyantBio公司簽署TissueofOrigin?(以下簡稱“TOO?”)全球權益和ZL轉讓協議,全資收購這款唯一獲FDA......

    這3個雜志撤回中國學者249篇文章,包含上交、中山等名校

    2021年7月20日,JournalofCellularPhysiology及JournalofCellularBiochemistry同時撤回了中國學者49篇文章。從2019年開始,Journalo......

    腫瘤治療的強心劑,中國學者開發腫瘤治療新策略

    磷酸甘油酸突變酶1(PGAM1)通過其代謝活性以及與其他蛋白質(例如α平滑肌肌動蛋白(ACTA2))的相互作用,在癌癥代謝和腫瘤進展中起關鍵作用。變構調節被認為是發現針對PGAM1的高選擇性和有效抑制......

    llumina宣布推出新型基因分型芯片|支持AllofUs研究計劃

    2018年12月6日,來自圣迭戈的消息——Illumina公司(納斯達克股票代碼:ILMN)今天宣布推出新型高密度基因分型芯片Infinium?GlobalDiversityArray。這款芯片設計源......

    SDSPAGE異常電泳現象及分析SDSPAGEHallofShame

    SDS-PAGE異常電泳現象及分析SDS-PAGEHallofShame.pdf  很不錯的東東~~推薦下~......

    Preparationofdenaturing6%

    Preparationofdenaturing6%polyacrylamidegelsformicrosatelliteanalysis(alsoforSSAP,high-resolutionIRAP......

  • <table id="4yyaw"><kbd id="4yyaw"></kbd></table>
  • <td id="4yyaw"></td>
  • 调性视频